System function | |
System | Automatic, Discrete, Random access, Bench-top with STAT sample priority |
Throughput | Up to 200 tests/hour |
Methodology | End-point, Fixed-time, Kinetic, Single/Dual reagent chemistries, Monochromatic/Bi-chromatic, Original system pack ready to use reagents |
Reagent/Sample Handling | |
Reagent/Sample tray | 30 positions for reagents and 30 positions for samples in 24-hour refrigerated compartment (2~12℃) |
Reagent volume | 10 - 350 µl, step by 0.5 µl |
Sample volume | 2 - 50 µl, step by 0.1 µl |
Reagent/Sample probe | Liquid level detection, vertical collision protection and inventory checking, reagent pre-warming |
Probe cleaning | Automatic interior and exterior washing |
Carryover | <3% |
Automatic sample dilution | Pre-dilution and post-dilution |
Internal bar code reader (optional) | |
Barcode Reader | Used for sample and reagent programming |
Capable to communicate with LIS in bi-directional mode | |
Reaction System | |
Reaction rotor | Rotating tray, containing 60 cuvettes |
Cuvette | Reusable, optical length 6 mm |
Reaction volume | 180 – 550 μl |
Operating temperature | 37 °C |
Temperature fluctuation | ± 0.2℃ |
Mixing Unit | |
Paddle Mixer | Independent mixing bar for homogenous mixing |
Laundry System | |
Cuvette Washing System | Washing station with pre-warmed detergent and water |
Upgraded auto-washing system ensures low carryover and low water consumption | |
7 step on-board washing system, Carryover <1% | |
Optical System | |
Light Source | Halogen-tungsten lamp |
Wavelength | 9 wavelengths, 340 nm, 405 nm, 450 nm, 510 nm, 546 nm, 578 nm, 620 nm, 670 nm & 700 nm |
Absorption range | 0 - 3.0 Abs, resolution 0.0001 Abs |
Detector | Silicon Photodiode |
Control and Calibration | |
Calibration modes | Linear (one point, two points and multi-points), Logit-Log 4P, Logit-Log 5P, spline, exponential, polynomial, parabola |
Control Rules | X-R, L-J, Westgard multi-rule, Cumulative sum check |
Operation Unit | |
Operation system | Windows 7, Windows 10 or latest |
Interface | RS-232 |
Working Conditions | |
Power Supply | 110/220 V, 50/60 Hz, 1000 VA or 100 – 130 V, 60 Hz, 1000 VA |
Water Consumption | 5 Litres/Hour |