<< Back to Film Thickness Analysis
Treatments are sometimes applied to improve the film quality of ultra-thin SiO₂ films deposited on Si substrates, and the UVISEL Plus can identify slight differences in film thickness and film quality before and after the treatment.
Through the treatment, a change in film thickness of several Å and a slight change in refractive index was confirmed . As a result of this treatment, the film thickness decreased, and an increase in refractive index suggests that the film became denser.
Spectroscopic Ellipsometer from FUV to NIR: 190 to 2100 nm