Product name | Water Distribution Monitor |
Model name | TW-100 |
Parameters | Turbidity, color, free chlorine, water pressure (standard); pH, conductivity, water temperature (option) |
Methods | See chart 1. |
Range | See chart 1. |
Repeatability | See chart 1. |
Display | LCD display: 320 x 240 dot B&W graphic LCD with backlight (touch panel type) Standard version simultaneously displays four measurement categories (turbidity, color, free chlorine, and water pressure); Optional version simultaneously displays up to seven parameters (the four parameters plus pH, conductivity, and water temperature) |
Calibration method | See chart 1. |
Automatic zero calibration | (Turbidity, color, free chlorine) Calibration method: Measurement of filtered sample water Trigger: Internal control (timer) or external control. Interval: 1 to 9999 hours (can be set as desired within that range) Time: 20~99 minutes(can be set as desired within that range) |
Automatic cleaning | (Turbidity, color) Cleaning method: Wiper cleaning Trigger: Internal control (timer) or external control. Interval: 5 to 9999 hours (can be set as desired within that range) (Free chlorine) Continuous cleaning with beads grinder |
Sample water conditions | Temperature: 0 to 40°C (must not freeze) Pressure: 0.1 to 0.75 Mpa Flow rate: over 100 mL/min Flow rate for measurement: 50 to 100 mL/min * Be sure to flush the sample water before starting the system. * Be sure to install a bypass for the pipe to which the system is attached. * If there is any possibility that the sample water could freeze, please protect the system from cold temperature. |
Ambient temperature/humidity | Temperature; 0 to 40°C; Humidity: 85% or less |
Analog output | Type: Measurement value: Same number as parameters (4 standard, 7 as maximum) Specifications: DC 4 to 20 mA, insulated output (no insulation between parameters), maximum load resistance 600 Ω |
Output signal | Type: Power off, batch alerts, maintenance Details: Power off: Output when power is off. Batch alerts: Cell temperature error, compensation temperature error, cell wiper error, internal communication error,water leak, battery error, upper and lower limit of measurement, light source error, zero calibration, span calibration,analyzer error Maintenance: Output when the system is in maintenance or calibration mode. Specifications: Non-voltage contact output, a contacts Contact ratings: AC 125 V, 0.3 A, DC 30 V, 1 A (depends on resistance load), each output COM is independent. |
Input signal | Type: Cleaning start, zero calibration start, error check sequence start, error water sampling start Details: Cleaning start: Cleaning procedure initiated with closed contact input Zero calibration start: Zero calibration initiated with closed contact input Error discrimination start: Error discrimination initiated with closed contact input Error water sample sampling start: Error water sample sampling initiated with closed contact input Specifications: Non-voltage contact input (open collector contact can be used), insulated input Resistance when ON: Maximum 100 W; Open voltage: DC 26 V; Short-circuit current: Max. 13 mA |
Communications | Interface: RS-232C compliant (*Remote monitoring system is option) |
Functions | Estimation of sample water volume (counting type), internal water leak detection, error water sampling |
Data memory | Measurement values for each parameters are stored in the unit’s data memory. Data can be transferred to a CF card. Memory interval: 1 minute or 1 hour Memory schedule: At the top of the hour Data memory longevity: 1 minute intervals: About 10 days; 1 hour intervals: About 1 year. * The latest data is written to up date a series of data. |
Wiring contact | Wiring ground: For power source: φ4.5 to φ6; For input/output: φ12.5 to φ14 |
Piping contact | Sample water input: Rc 1/4; Drainage: Rc 1/4; Calibration solution input: Rc 1/8; Error sample sampling aperture: Rc 1/8 |
Construction | Designed for indoor installation. * Housed in a special case (frame) for exterior installation (option). |
Power | AC 100 V to AC 230 V, ±10%, 50/60 Hz |
Power consumption | AC 100 V to AC 120 V: 120 VA; AC 100 V to AC 230 V: 150 VA |
Mass | Approx. 15 kg (Measurement unit only) |
External dimensions | 350 mm (W) x 160 mm (D) x 420 mm (H) (not including installation fixtures and projecting portions) |
Paint color/Material | Munsell 5PB8/1; Unit case is made of SUS. |
Installation environment | Install on a flat surface without vibration shocks. No dust, mist, or corrosive gases in the atmosphere. Lower air pressure. No direct sunlight. Ventilated air. 2,000 m or less above sea level. |