Model | CS-151F1 | CS-152F1 | CS-150F1 |
Target of measurement | SC-1 (ammonia/hydrogen peroxide aqueous solution) | SC-2 (hydrochloric acid/hydrogen peroxide aqueous solution) | SPM (sulfuric acid/hydrogen peroxide aqueous solution) |
Multi-monitoring with one unit of up to 4 types of chemical solutions or ranges possible. | |||
Measurement principle | Absorption spectrometry | ||
Concentration computation principle | Temperature-compensated multivariate analysis | ||
Measurement range (mass %)
| NH3: 0.00% to 1.00% H2O2: 0.00% to 5.00% H2O : 94.0% to 100.0% | HCI : 0.00% to 2.00% H2O2: 0.00% to 2.00% H2O : 96.0% to 100.0% | H2SO4: 70.0% to 96.0% H2O2: 0.00% to 10.00% H2O : 4.0% to 30.0% |
Repeatability (mass %) (within ±1% room temperature)
| NH3: ±0.15% H2O2: ±0.30% H2O : ±1.5% | HCI : ±0.15% H2O2: ±0.15 % H2O : ±1.5% | H2SO4: ±0.50% H2O2: ±0.50% H2O : ±1.5% |
Sample solution temperature | 20 to 80 : (when using Inline cell) 20 to 30 : (when using cell unit) | ||
Measurement cycle | Approx. 3 seconds | ||
Content of | • 1st component: measured concentration value (%) | • Error code |
|
LCD and input
| • 2nd component: measured concentration value (%) • 3nd component: measured concentration value (%) • 1st and 2nd components: concentration alarm status | • Parallel concentration alarm output invalid status • Serial/parallel status • Background compensation data |
|
RS-232C serial input | • Serial/parallel status switching • Measurement data transfer request • 1st and 2nd component concentration alarm setting | • Parallel concentration alarm output status (valid/invalid) • Target measurement solution switching |
|
Parallel input | 12 to 30 V DC (photocoupler insulated) | • Parallel concentration alarm output • Target measurement solution switching • Background compensation |
|
Parallel output | Open collector output (photocoupler insulated) Max. current when on: DC 5 mA (no internal protection resistor) Max. ignition voltage when off: 30 V DC
| • 1st component concentration alarm, 4 types • 2nd component concentration alarm, 4 types • Device error • Measuring • Target measurement solution type • Background compensation status |
|
Solution temperature input | Terminal block connection (M3 terminal screws) | ||
Analog output | • 1st component concentration alarm, 4 types: 4 to 20 mA • 2nd component concentration alarm, 4 types: 4 to 20 mA * We also support other measurement ranges. Please specify when ordering. | ||
Fiber optic cable | Dedicated CS-100F1 Series fiber optic cable (25 ±15: installation temperature) | ||
ample cell | We can supply dedicated sample cells (inline cell or cell unit) for various kinds of chemical solutions. | ||
Power supply | 100 V to 230 V AC (single phase) ±10%, 50/60 Hz, approx. 100 VA | ||
Dimensions | 205 (W) x 329 (D) x 269 (H) mm 8.1 (W) x 13.0 (D) x 10.6 (H) in (excluding protruding part) | ||
Weight | Approx. 11 kg | ||
Analog output | 4 to 20 mA (2 outputs: H2SO4: 50.0 to 100.0 %/H2O: 0.00 to 10.00 %) The analog output can be prepared to correspond to each requirement by users. (Please specify when ordering.) | ||
Power | 24 V DC ±10%, 2 A | ||
Dimensions | 205 (W) x 329 (D) x 269 (H) mm 8.1 (W) x 13 (D) x 10.6 (H) in | 291 (W) x 329 (D) x 269 (H) mm 11.5 (W) x 13 (D) x 10.6 (H) in | |
Mass | Approx. 11 kg | Approx. 12 kg | |
Connection pipe size | For inlet and outlet: Outer diameter 6 mm (inner diameter 4 mm) | ||
Pipe connection | Nippon Pillar Super type |