Model | CS-152 | CS-152C | |
Model specifications | No cooling unit | Integrated cooling unit | |
Measurement sample | SC-2 (Hydrochloric acid/Hydrogen peroxide aqueous solution) | ||
Measurement principle | Absorption spectrometry | ||
Concentration computation principle | Temperature-compensated multivariate analysis | ||
Measurement range | HCl: 0.00 to 2.00 %/0.00 to 10.00 % | ||
Reproducibility | HCl: ±0.15 %/±0.20 % | ||
(Peak to peak)
| H2O2: ±0.15 %/±0.20 % | ||
Chemical conditions
| Temperature | 20 to 30 °C | 20 to 80 °C |
Flow rate | 20 to 60 ml/min. | 20 to 40 ml/min. | |
Pressure | 0.10 to 0.20 MPa | ||
Measurement cycle | Approx. 3 seconds minimum | ||
Equipment temperature | 20 to 25 °C | ||
Operation panel | 1)Temperature level display: 2)Alarm display (Hydrochloric acid, Hydrogen peroxide): 3)Alarm value settings (Hydrochloric acid, Hydrogen peroxide): | ||
Serial input/output | RS-232C 2)Alarm value settings (Hydrochloric acid, Hydrogen peroxide): | ||
Parallel input | 12 to 30 V DC (photocoupler insulated) | ||
Parallel output | Maximum current when open collector output is ON: 5 mA 1)Alarm output (Hydrochloric acid, Hydrogen peroxide): 2)Device error (including liquid leakage) 3)Under measurement | ||
Analog output | 4 to 20 mA (2 outputs: HCl concentration, H2O concentration | ||
Power | 24 V DC ±10%, 2 A | ||
Dimensions | 205 (W) x 329 (D) x 269 (H) mm | 291 (W) x 329 (D) x 269 (H) mm | |
Mass | Approx. 11 kg | Approx. 12 kg | |
Connection pipe size | For inlet and outlet: Outer diameter 3 mm (inner diameter 2 mm) | ||
Pipe connection | Nippon Pillar Super type | ||
Utilities | Instrumentation air for driving air valve: |