HP-300

現場安裝型pH計

HORIBA 的長期暢銷水質計在室內和室外應用中均表現出色。

適用於各種生產過程中的pH監測與控制,以及廢水處理等應用。我們提供各種電極和清潔設備來適應各種樣品。

產品分類: Process Water
製造商: HORIBA Advanced Techno Co., Ltd.

Lineup of a wide range of electrodes

A wide variety of electrodes are offered to cater to different needs.

Comprehensive range of cleaning devices.

A diverse lineup of cleaning devices are offered.  Particularly, ultrasonic cleaners are advantageous as they do not require water or air, making them convenient for retrofitting existing HP-300 installations on site. The drop-in type of cleaning device, which fits into a holder, offers ease of maintenance.

Product namepH meter
ModelHP-300
Combined
sensor
Glass electrode
Measurable rangepHpH0 to pH14 (display range: pH−1 to pH15)
Temperature0℃ to 100℃
Display range when automatic detection capability for each temperature
sensor type is used: −10℃ to 110℃
Display range when temperature sensor type is manually set: −20℃ to 130℃
Display
resolution
pH0.01 pH
Temperature0.1℃
PerformancepHRepeatabilityWithin ±0.03 pH (response for equivalent input)
LinearityWithin ±0.03 pH (response for equivalent input)
TemperatureRepeatability±0.3℃ (response for equivalent input)
Linearity±0.3℃ (response for equivalent input)
Transmission
output
Output type4 mA to 20 mA DC: input/output isolated type
Load resistanceMaximum: 600Ω Case of 24 V DC power supply (*Note 1)
LinearityWithin ±0.08 mA (output only)
RepeatabilityWithin ±0.08 mA (output only)
Output rangepH: Selection from preset ranges or free range input within measuring range.
Occasional out for errorHold or burnout to either 3.8 mA or 21 mA
Transmission holdIn the maintenance mode, transmission signal is held at the latest value or preset value.
In the calibration mode, transmission signal can be alive or held.
Contact inputNumber of input points1
Contact typeNo-voltage "a" contact for open collector
ConditionsON resistance: 40 Ω 
Open voltage: 1.2 V
Short-circuit current: 21 mA DC max.
Contact functionExternal input for transmission holding.
Temperature
compensation
Applicable temperature
element
Platinum resistor: 1 kΩ (0℃)
Positive relation resistor with temperature
:500 Ω (25℃), 6.8 kΩ (25℃), 10 kΩ (25℃)
Element selection methodAutomatic temperature sensor type detection or manual selection
(No temperature element is selectable)
Temperature compensation
range
0℃ to 100℃
Temperature calibration1 point calibration comparing reference thermometer
CalibrationCalibration methodAuto or basic (manual) calibration
Number of calibration pointsSelectable from 1, 2, and 3 points
Kinds of standard solutionspH 2, 4, 7, 9, 10
Any pH standard solution (with difference of 2 pH or more) for basic calibration
Additional capabilitiesAutomatic detection of kind of standard solution
Automatic detection of electric potential stability
Automatic detection of calibration failure
(asymmetry potential, sensitivity, or response time)
Calibration history (asymmetry potential, sensitivity, and number of days elapsed after last calibration)
Self-checkCalibration errorAsymmetry potential error, sensitivity error, and response time error
Temperature calibration error
Standard solution detection error
Electrode diagnostic errorCracking of glass response membrane
Reference electrode impedance error (only applicable for differential circuit mode with a liquid ground electrode)
Temperature sensor short-circuit, temperature sensor disconnection, and deviation from temperature measurement range
Meter errorCPU, ADC, and EEPROM errors
Operating temperature range−20℃ to 60℃ (without freeze)
Operating humidity rangeRelative humidity: 5% to 90% (without condensation)
Storage temperature−25℃ to 65℃
Power supplyRated voltage24 V DC (operating voltage range: 21 V to 32 V DC) (*Note 1)
Power consumption0.6 W max.
Compatible
standards
CE markingCompatible
standards
EMC directiveEN61326-1
RoHS DirectiveEN50581
ImmunityIndustrial electromagnetic environment
Electostatic dischargeIEC61000-4-2
Electromagnetic field of radiated
radio frequency
IEC61000-4-3 (*Note 2)
Electric fast transient/burstIEC61000-4-4 (*Note 2)
SurgeIEC61000-4-5 (*Note 3)
Conducted interference induced
by radio frequency
IEC61000-4-6 (*Note 2)
EmissionClass A
Radiated disturbanceCICPR 11 CLASS A
FCC rulesPart15 CLASS A
StructureInstallationOutdoor installation type
Installation methodMount on 50 A pole or wall
Protection codeIP65IEC60529, JIS C0920
Case materialAluminum alloy (coated with epoxy-denatured melamine resin)
Material of fittingsSUS304
Material of hoodSUS304 stainless steel (coated with epoxy-denatured melamine resin)
Material of windowPolycarbonate
Display elementReflective monochrome LCD
External dimensions180 (W) mm × 155 (H) mm × 115 (D) mm (excluding mounting brackets)
MassMain body: Approx. 2.8 kg
cover and mounting brackets: Approx. 1 kg

Note

  • Note 1:  The maximum load resistance is available within the range shown in the figure below, depending on power supply voltage:

 

Relation between power supply voltage and load resistance

Relation between power supply voltage and load resistance

  • Note 2:  The standard for effect on the reading by the electromagnetic field of the radiated radio frequency and by the conducted interference is within the measured pH value 0.1 pH.
  • Note 3:  When the sensor cable, the transmission cable, or the contact input cable is extended by 30 m or more, the surge test under the EMC Directive for CE marking is not applied.
  • Note 4:  An arrester (spark over voltage: 400 V) is implemented for transmission output, contact input, and communication. However, use a most suitable surge absorption element on the connection lines in accordance with the ambient environment, the situation of equipment installed, and the externally connected equipment.

留言諮詢

如您有任何疑問,请在此留下詳細需求或問題,我們將竭誠您服務。

* 這些欄位為必填項目。

更多推荐

Glass Electrode 1076A-10C
Glass Electrode 1076A-10C

For measurement of low-conductivity water and non-aqueous solvents.

H-1 Series
H-1 Series

Field-installation type water quality measuring instruments

HD-960LR
HD-960LR

Dissolved Oxygen Concentration Monitor Series for Semiconductor Manufacturing

LAQUA DO210
LAQUA DO210

Handheld Dissolved Oxygen/Temperature Meter

LAQUA DO220
LAQUA DO220

Handheld Dissolved Oxygen/Temperature Meter

LAQUA DS-72
LAQUA DS-72

Benchtop Conductivity/Resistivity/Total Dissolved Solids/Salinity/Temperature Meter

LAQUA EC1100
LAQUA EC1100

Benchtop Water Quality Meters

LAQUA EC210
LAQUA EC210

Handheld Conductivity/Resistivity/Salinity/Total Dissolved Solids/Temperature Meter

LAQUA EC220
LAQUA EC220

Handheld Conductivity/Resistivity/Salinity/Total Dissolved Solids/Temperature Meter

LAQUA F-72
LAQUA F-72

Benchtop pH/ORP/Ion/Temperature Meter

LAQUA F-73
LAQUA F-73

Benchtop pH/ORP/Ion/Temperature Meter

LAQUA F-74
LAQUA F-74

Benchtop pH/ORP/Ion/Conductivity/Resistivity/Total Dissolved Solids/Salinity/Temperature Meter

LAQUA PC1100
LAQUA PC1100

Benchtop Water Quality Meters

LAQUA PC210
LAQUA PC210

Handheld pH/ORP/Conductivity/Resistivity/Total Dissolved Solids/Salinity/Temperature Meter

LAQUA PC220
LAQUA PC220

Handheld pH/ORP/Conductivity/Resistivity/Total Dissolved Solids/Salinity/Temperature Meter

LAQUA PD2000
LAQUA PD2000

Benchtop pH/ORP/Ion/DO/BOD/Temperature Meter

LAQUA PD210
LAQUA PD210

Handheld pH/ORP/Dissolved Oxygen/Temperature Meter

LAQUA PD220
LAQUA PD220

Handheld pH/ORP/Dissolved Oxygen/Temperature Meter

LAQUA PH1100
LAQUA PH1100

Benchtop Water Quality Meters

LAQUA PH1200
LAQUA PH1200

Benchtop Water Quality Meters

LAQUA PH1300
LAQUA PH1300

Benchtop Water Quality Meters

LAQUA PH1500
LAQUA PH1500

Benchtop Water Quality Meters

514-05
514-05

0.5% NaCl Standard Solution

514-22
514-22

1.41 mS/cm Conductivity Standard Solution

514-23
514-23

12.9 mS/cm Conductivity Standard Solution

514-50
514-50

5.0% NaCl Standard Solution